PROTEKSI RADIASI RUANG RADIOLOGI DI UPTD RSUD DATU BERU KABUPATEN ACEH TENGAH
Pengarang
Rizke Dian Oktari - Personal Name;
Dosen Pembimbing
Rini Safitri - 197004251995122001 - Dosen Pembimbing I
Nomor Pokok Mahasiswa
1608102010015
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Penerbit
Banda Aceh : Fakultas MIPA (S1)., 2022
Bahasa
Indonesia
No Classification
539.2
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Proteksi radiasi dilakukan untuk mengurangi pengaruh radiasi akibat paparan radiasi, proteksi radiasi menjadi suatu keharusan agar terwujudnya keselamatan radiasi. Adapun rumusan masalah dalam penelitian ini adalah mengetahui proteksi radiasi pada ruangan radiologi di RSUD Datu Beru dan mengetahui dosis yang diterima radiografer dengan tujuan untuk menganalisis proteksi radiasi pada ruangan radiologi RSUD Datu Beru, mengukur sebaran dosis radiasi dibalik shielding ruang pemeriksaan sinar-X dan untuk membandingkan dosis yang diterima radiografer dengan Nilai Batas Dosis (NBD). Metode yang digunakan adalah melabeli dinding (baik dengan huruf atau angka) untuk membedakan jarak antara sumber radiasi dengan dinding ruangan. Diukur pula beban kerja baik pada radiografer maupun masyarakat umum untuk melihat pemenuhan standar Nilai Batas Dosis (NDB). Alat yang digunakan surveymeter merk radex RD1503+, nilai yang keluar dari layar alat surveymeter dicatat kemudian dihitung untuk mengetahui ada tidaknya kebocoran ruangan. Hasil pengukuran sebaran dosis radiasi dibalik dinding penahan radiasi (shielding) dengan surveymeter adalah ±0,80 μSv/jam atau 1,152 mSv/tahun atau 0,288 mSv per tiga bulan yang menunjukkan adanya kebocoran didinding ruang pemeriksaan sinar-X. Dosis yang diterima pekerja radiasi (radiografer) dengan data yang sudah ada pada 5 orang pekerja radiasi adalah paling rendah 0,009 mSv per tiga bulan dan paling tinggi 0,024 mSv per tiga bulan.
Kata Kunci : dosis hambur, proteksi, proteksi radiasi, radiasi, surveymeter
Radiation protection is carried out to reduce the influence of radiation due to radiation exposure, radiation protection is a must for the realization of radiation safety. The formulation of the problem in this study is to determine the dose received by the radiographer with the aim of analyzing radiation protection in the radiology room of Datu Beru Hospital, measuring radiation dose spread behind X-ray examination room shield and to compare the dose received by the radiographer with the dose limit value (NBD). The method used is to label the walls (either with letters or numbers) to distinguish the distance between the radiation source and the walls of the room, the workload of both radiographers and the general public is also measured to see if the dose limit value (NBD) is met. The tool used is a surveymeter brand radex RD1503+, the value that comes out of the surveymeter tool screen is recorded and then calculated to find out whether there is a room leak or not. The result of measuring the dose distribution behind the shielding with a surveymeter is ±0,80 μSv/hour or 1,152 mSv/year or 2,088 mSv/3 months which indicates a leak in the walls of the X-ray examination room. The dose received by the radiographer with available data on 5 radiographers, the lowest was 0,009 mSv/3 months and the highest was 0,024 mSv/3 months or equivalent to 50 mSv/year. Keywords : protection, radiation, radiation protection, scattering dose, surveymeter
STUDI PROTEKSI RADIASI SINAR-X TERHADAP PEKERJA RADIASI DI RUMAH SAKIT UMUM DAERAH KELAS B SIGLI (MAULIA SARI, 2020)
HUBUNGAN ANTARA HEALTH BELIEF DENGAN PERILAKU COMPLIANCE PADA PASIEN GAGAL GINJAL KRONIS YANG MENJALANI HEMODIALISA DI UPTD RSUD DATU BERU KABUPATEN ACEH TENGAH (DINA QISTHI, 2023)
ANALISIS PENAHAN RADIASI PESAWAT SINAR–X RUANG RADIOGRAFI UMUM SEBAGAI UPAYA PROTEKSI RADIASI DI RSUD DR. ZAINOEL ABIDIN BANDA ACEH (Muliana Z, 2024)
PENGARUH MARKETING MIX TERHADAP RETENSI PASIEN DI RUMAH SAKIT UMUM DAERAH (RSUD) DATU BERU TAKENGON (Rizki Anjar Pinanggih, 2022)
SURVEI KESESUAIAN RUANG RADIOLOGI DI RUMAH SAKIT UMUM DAERAH LANGSA SEBAGAI UPAYA UNTUK PROTEKSI RADIASI (Putri Mawaddatul Muna, 2019)