UJI KESESUAIAN PESAWAT SINAR X FLUOROSKOPI DI RUMAH SAKIT TYPE A BANDA ACEH
Pengarang
HUSNUL HAYATI - Personal Name;
Dosen Pembimbing
Nomor Pokok Mahasiswa
1608102010032
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : Universitas Syiah Kuala., 2021
Bahasa
Indonesia
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
ABSTRAK
Telah dilakukan penelitian uji kesesuaian pesawat sinar-X fluoroskopi di rumah sakit type A Banda Aceh. Penelitian ini bertujuan untuk mengetahui akurasi tegangan, kualitas berkas radiasi (HVL), laju dosis maksimum diudara dan laju dosis tipikal sebagai uji kualitas berkas sinar-X pada pesawat fluoroskopi di rumah sakit type A di Banda Aceh. Pengukuran ini menggunakan Multimeter X- ray dan fantom PMMA. Alat yang digunakan dalam penelitian ini adalah pesawat fluoroskopi dan filter Alumunium dengan ketebalan 1,5 mm dan 0,5 mm. Hasil penelitian menunjukkan bahwa akurasi tegangan yang didapatkan pada fluoroskopi dalam rentang 0,1- 0,3%, hasil ini masih sesuai dengan ambang batas maksimum yang diperkenankan oleh BAPETEN yakni kVp ? 10%. Nilai kualitas berkas radiasi (HVL) yang diperoleh lebih kecil dari nilai HVL minimum yang ditetapkan BAPETEN artinya, kualitas berkas pesawat fluoroskopi berada dibawah nilai ambang yang diperkenankan oleh BAPETEN. Hasil laju dosis maksimum di udara pada pesawat fluoroskopi masih berada di bawah nilai batas ambang yang ditetapkan BAPETEN, kecuali pada tegangan 100 kVp sebesar 60,48 mGy/menit. Hasil laju dosis tipikal pada pesawat fluoroskopi masih memenuhi nilai ambang yang ditetapkan oleh BAPETEN.
Kata Kunci : fluoroskopi, uji kesesuaian, akurasi tegangan, laju dosis, kualitas berkas radiasi
ABSTRACT
A study on quality control in X-ray fluoroscopy has been carried out in the Banda Aceh type A hospital. This study aims to determine the accuracy of the voltage, the quality of the radiation beam (HVL), the maximum dose rate in the air and the typical dose rate as a test of the quality of the X-ray beam on a fluoroscopy in a type A hospital in Banda Aceh. This measurement uses an X ray Multimeter and a PMMA phantom. The tools used in this study were fluoroscopy and aluminum filters with a thickness of 1.5 mm and 0.5 mm. The results showed that the voltage accuracy obtained in fluoroscopy was in the range of 0.1- 0.3%, this result was still in accordance with the threshold allowed by BAPETEN, namely kVp ? 10%. The value of the radiation beam quality (HVL) obtained is smaller than the minimum HVL value set by BAPETEN, which means, the conclusion of the maximum dose rate in the air on the X-ray fluoroscopy are still below the threshold value set by BAPETEN, except at a voltage of 100 kVp of 60.48 mGy / minute. The results of a typical dose rate on a fluoroscopy still meet the threshold value set by BAPETEN.
Keywords : fluoroscopy, quality control, voltage accuracy, dose rate, radiation beam quality
Tidak Tersedia Deskripsi
ANALISIS UJI KESESUAIAN GENERATOR DAN TABUNG PESAWAT SINAR-X DI RUANG RADIOLOGI RSUD ACEH BESAR (DITA SAFIRA, 2023)
UJI KESESUAIAN PESAWAT SINAR-X MEREK MINDRAY DI INSTALASI RUANG RADIOLOGI RUMAH SAKIT UMUM MEURAXA (Egidia Tesalonika, 2024)
QUALITY COTROL (KENDALI MUTU) PESAWAT SINAR-X MERK SIEMENS HELTINERS DI INSTALASI RADIOLOGI RUMAH SAKIT PENDIDIKAN UNIVERSITAS SYIAH KUALA (Mauliana, 2024)
UJI KESESUAIAN PADA PESAWAT MAMMOGRAFI DI INSTALASI RADIOLOGI RUMAH SAKIT UMUM DAERAH DR. ZAINOEL ABIDIN (DITA DUHITA, 2021)
UJI KESESUAIAN GENERATOR DAN TABUNG PESAWAT SINAR-X DI LABORATORIUM KLINIK FAKULTAS KEDOKTERAN HEWAN UNIVERSITAS SYIAH KUALA (alya rizka maiza, 2023)