UJI KENDALI MUTU CITRA CT SCAN 64 SLICE BERDASARKAN NILAI CT NUMBER
Pengarang
Yulli Malina - Personal Name;
Dosen Pembimbing
Elin Yusibani - 197903232006042002 - Dosen Pembimbing I
Evi Yufita - 197509202000122001 - Dosen Pembimbing III
Nomor Pokok Mahasiswa
2008102010045
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : Fakultas MIPA Fisika., 2025
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
ABSTRAK
Kajian tentang uji kendali mutu citra CT scan 64 slice berdasarkan nilai hounsfield unit (Hu) menggunakan dua Phantom air dengan tahun keluaran yang berbeda yaitu 2008 dan 2018 yang dilakukan di Rumah Sakit dr. Zainoel Abidin, Banda Aceh. Penelitian ini dilakukan bertujuan untuk megetahui kualitas citra dari CT scan 64 slice dengan melakukan analisis terhadap CT number, Keseragaman Noise, dan Keseragaman pusat dan tepi menggunakan aplikasi ImageJ dan dilakukan perbandinggan dengan standar PARKA BAPETEN terhadap uji kualitas citra CT scan. Pada penelitian ini menggunakan dua phantom dengan tahun yang berbeda yaitu 2008 dan 2018 dimana dari masing-masing phantom menggunakan tiga citra. Hasil analisis penelitian menunjukkan bahwa nilai CT number dari phantom 2008 berada pada 0,2 Hu hingga 1,1 Hu dan pada phantom 2018 berada pada -1,1 Hu hingga -1,4 Hu dimana nilai-nilai tersebut berada pada batas toleransi PARKA BAPETEN untuk CT number yaitu ±4 Hu. Pada hasil analisis keseragaman noise untuk phantom 2008 berada pada 0,2 Hu hingga 0,6 Hu dan untuk phantom 2018 berada pada 0,2 Hu hingga 0,3 Hu dimana nilai-nilai tersebut berada pada batas toleransi PARKA BAPETEN untuk keseragaman noise yaitu ≤2 Hu. Pada hasil analisis keseragaman pusat dan tepi untuk phantom 2008 berada pada -1.2 Hu hingga -2.9 Hu dan untuk phantom 2018 berada pada -1,2 Hu hingga -1,7 Hu dimana nilai-nilai tersebut berada pada batas toleransi PARKA BAPETEN untuk keseragaman pusat dan tepi yaitu ≤2 Hu dari hasil penelitian tersebut dapat disimpulkan bahwasanya pesawat CT scan 64 slice dirumah sakit tipe A Banda Aceh untuk uji kualitas citra masih sesuai dengan standar PARKA BAPETEN.
Kata Kunci : CT Number, Keseragaman Noise, Keseragaman Pusat dan Tepi
ABSTRACT The study on the quality control testing of 64-slice CT scan images based on the Hounsfield Unit (Hu) values using two different air phantoms from different production years (2008 and 2018) was conducted at Dr. Zainoel Abidin Hospital, Banda Aceh. The purpose of this study was to evaluate the image quality of the 64-slice CT scan by analyzing the CT number, noise uniformity, and center-edge uniformity using the ImageJ application, and comparing the results with the PARKA BAPETEN standards for CT image quality testing. In this study, two phantoms from different years (2008 and 2018) were used, with three images taken from each phantom. The results of the analysis showed that the CT number for the 2008 phantom ranged from 0,2 to 1,1 Hu, while the 2018 phantom ranged from -1,1 Hu to -1,4 Hu. These values fall within the tolerance limits of the PARKA BAPETEN standard for CT numbers, which is ±4 Hu. The noise uniformity analysis showed values ranging from 0,2 Hu to 0,6 Hu for the 2008 phantom and from 0,2 Hu to 0,3 Hu for the 2018 phantom. These values are within the PARKA BAPETEN tolerance limit for noise uniformity, which is ≤2 Hu. For the center-edge uniformity analysis, the 2008 phantom showed values between -1,2 Hu and -2,9 Hu, while the 2018 phantom ranged from -1,2 Hu to -1.,7 Hu. These values also fall within the PARKA BAPETEN tolerance limit for center-edge uniformity, which is ≤2 Hu. Based on the findings of this study, it can be concluded that the 64-slice CT scanner at Dr. Zainoel Abidin Hospital in Banda Aceh complies with the PARKA BAPETEN standards for image quality testing. Keywords: CT Number, Noise Uniformity, Center and Edge Uniformity
ANALISIS PENGUKURAN COMPUTED TOMOGRAPHY DOSE INDEX (CTDI) DENGAN MENGGUNAKAN KAMAR IONISASI (IONIZATION CHAMBER) PADA PEMERIKSAAN KEPALA (Ruhul Jannati, 2013)
ESTIMASI DOSIS RADIASI PADA PEMERIKSAAN CT- SCAN TORAKS BERDASARKAN FAKTOR EKSPOSI DAN KUALITAS CITRA MENGGUNAKAN PARAMETER PENILAIAN SIGNAL TO NOISE RATIO (SNR) (Ayunda Fajarni Rahmi, 2024)
PENGUKURAN DIAMETER ABDOMEN PASIEN DEWASA TERHADAP ARUS TABUNG DENGAN METODE COMPUTED TOMOGRAPHY DOSE INDEX (CTDI) MENGGUNAKAN PESAWAT CT-SCAN (PUTRI SAFIRA, 2018)
ANALISIS HASIL CITRA CT-SCAN ORGAN TUBUH DENGAN MEDIA KONTRAS IOPAMIDOL 370 (KHAIRUNNISA, 2022)
TEKNIK SEGMENTASI BERBASIS ACTIVE CONTOUR UNTUK PENGUKURAN DIAMETER BATU GINJAL MENGGUNAKAN CITRA COMPUTED TOMOGRAPHY SCAN (CT-SCAN) (Yanti Armayani, 2019)