QUALITY COTROL (KENDALI MUTU) PESAWAT SINAR-X MERK SIEMENS HELTINERS DI INSTALASI RADIOLOGI RUMAH SAKIT PENDIDIKAN UNIVERSITAS SYIAH KUALA
Pengarang
Mauliana - Personal Name;
Dosen Pembimbing
Rini Safitri - 197004251995122001 - Dosen Pembimbing I
Evi Yufita - 197509202000122001 - Dosen Pembimbing II
Nomor Pokok Mahasiswa
1908102010018
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : ., 2024
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Rumah Sakit Pendidikan Universutas Syiah Kuala memiliki fasilitas ruang radiologi dengan mobile X-ray yang digunakan untuk keperluan diagnosa penyakit, dan alat tersebut belum pernah dilakukan program kendali mutu (quality control). Quality control merupakan program jaminan mutu yang didalamnya termasuk uji kesesuaian. Uji kesesuaian adalah pengujian yang dilakukan untuk memastikan pesawat sinar-X dalam kondisi yang andal atau layak dan memenuhi peraturan perundang-undangan yang ditetapkan. Uji kesesuaian bertujuan untuk keamanan pasien, pekerja dan masyarakat dalam sistem pengoperasian pesawat sinar-X dengan baik dan bena. Uji yang dilakukan dalam penelitian adalah uji akurasi tegangan, uji linieritas dan uji berkas kualitas sinar-X melalui pengujian HVL (Half Value Layer). Pengujian ini dilakukan dengan menggunakan multimeter X-ray. Hasil pengukuran akan dibandingkan dengan perka BAPETEN nomor 2 tahun 2022, dimana uji akurasi tegangan tegangan ± ≤ 10%, uji linieritas ± ≤ 0,1 dan uji HVL ≥ 2,1 mmAl untuk 70 kV dan ≥ 2,3 mmAl untuk 81 kV. Hasil yang diperoleh pada penelitian ini untuk pesawat sinar-X merk Siemens Heltiners adalah uji akurasi tegangan memiliki nilai penyimpangan terbesar adalah 2,42% pada 70 kV. Uji linieritas untuk fokus kecil sebesar 0,035 dan untuk fokus besar 0,094 Pengukuran nilai HVL adalah 2,99 mmAL pada 70 kV serta 3,36 mmAL pada 81 kV. Berdasarkan hasil yang diperoleh untuk masing-masing uji kesesuaian masih berada dibawah batas aman yang telah ditetapkan oleh BAPETEN, sehingga dapat disimpulkan bahwa pesawat sinar-X di ruang radiologi rumah sakit Pendidikan Universitas Syiah Kuala dalam kondisi yang andal sesuai Perka BAPETEN nomor 2 tahun 2022.
Kata kunci : Uji Kesesuaian, Sinar-X, Multimeter X-ray, mobile x-Ray
The Syiah Kuala University Teaching Hospital has radiology room facilities with mobile X-ray which are used for disease diagnosis purposes, and this tool has never had a quality control program. Quality control is a quality assurance program which includes conformity testing. Conformity tests are tests carried out to ensure that the X-ray aircraft is in a reliable or proper condition and meets established statutory regulations. The conformity test aims to ensure the safety of patients, workers and the public in the proper and correct operation of the X-ray aircraft system. The tests carried out in the research were voltage accuracy tests, linearity tests and X-ray beam quality tests through HVL (Half Value Layer) testing. This test is carried out using an X-ray multimeter. The measurement results will be compared with BAPETEN's regulation number 2 of 2022, where the voltage accuracy test is ± ≤ 10%, the linearity test is ± ≤ 0.1 and the HVL test is ≥ 2.1 mmAl for 70 kV and ≥ 2.3 mmAl for 81 kV. The results obtained in this research for the Siemens Heltiners brand X-ray aircraft were that the voltage accuracy test had the largest deviation value of 2.42% at 70 kV. The linearity test for a small focus is 0.035 and for a large focus 0.094. The HVL value measurement is 2.99 mmAL at 70 kV and 3.36 mmAL at 81 kV. Based on the results obtained for each conformity test, it is still below the safe limits set by BAPETEN, so it can be concluded that the X-ray aircraft in the radiology room at the Syiah Kuala University Teaching Hospital is in reliable condition according to BAPETEN Perka number 2 of 2022. Keywords: Conformity Test, X-Ray, X-ray Multimeter, mobile x-Ray
SURVEI KESESUAIAN RUANG RADIOLOGI DI RUMAH SAKIT UMUM DAERAH LANGSA SEBAGAI UPAYA UNTUK PROTEKSI RADIASI (Putri Mawaddatul Muna, 2019)
UJI KESESUAIAN PESAWAT SINAR-X MEREK MINDRAY DI INSTALASI RUANG RADIOLOGI RUMAH SAKIT UMUM MEURAXA (Egidia Tesalonika, 2024)
UJI KESESUAIAN PADA PESAWAT MAMMOGRAFI DI INSTALASI RADIOLOGI RUMAH SAKIT UMUM DAERAH DR. ZAINOEL ABIDIN (DITA DUHITA, 2021)
ANALISIS UJI KESESUAIAN GENERATOR DAN TABUNG PESAWAT SINAR-X DI RUANG RADIOLOGI RSUD ACEH BESAR (DITA SAFIRA, 2023)
PEMANTAUAN PROTEKSI RADIASI SINAR-X DI RUANG RADIOLOGI RUMAH SAKIT UMUM RUJUKAN PEMERINTAH ACEH UTARA (WAHYUDDIN, 2022)