ANALISIS PENAHAN RADIASI PESAWAT SINAR–X RUANG RADIOGRAFI UMUM SEBAGAI UPAYA PROTEKSI RADIASI DI RSUD DR. ZAINOEL ABIDIN BANDA ACEH
Pengarang
Muliana Z - Personal Name;
Dosen Pembimbing
Rini Safitri - 197004251995122001 - Dosen Pembimbing I
Evi Yufita - 197509202000122001 - Dosen Pembimbing II
Nomor Pokok Mahasiswa
1908102010024
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : Fakultas MIPA Fisika., 2024
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Pemantauan paparan radiasi pada setiap daerah kerja perlu dilakukan untuk memastikan bahwa penahan radiasi yaitu dinding pada ruang penyinaran memenuhi persyaratan keselamatan radiasi. Penelitian ini bertujuan untuk mengevaluasi nilai faktor transmisi untuk penahan radiasi sesuai dengan ketentuan NCRP 147, dan untuk mengetahui hasil pengukuran laju paparan radiasi di sekitar ruang radiografi umum pada Instalasi Radiologi RSUD dr. Zainoel Abidin Banda Aceh. Metode penelitian yang dilakukan adalah menghitung ketebalan dinding penahan radiasi dengan menggunakan ketentuan dari NCRP 147, kemudian membandingkan hasil perhitungan penahan radiasi dengan ketebalan dinding yang sudah dibangun, selanjutnya melakukan pengukuran laju paparan radiasi di titik-titik yang sudah ditentukan. Hasil perhitungan penahan radiasi yang terpasang di ruang radiografi umum pada Instalasi Radiologi RSUD dr. Zainoel Abidin Banda Aceh dengan tebal 30 cm beton dan 2 mm Pb, untuk semua dinding memenuhi persyaratan dan aman untuk pekerja radiasi dan masyarakat umum. Hasil perhitungan paparan radiasi di ruang radiografi umum pada Instalasi Radiologi RSUD dr. Zainoel Abidin Banda Aceh diperoleh nilai yang jauh dibawah nilai batas dosis yang telah ditetapkan sehingga nilai dosis masih dalam kategori aman.
Kata kunci: Sinar-X, Proteksi Radiasi, Penahan Radiasi, Laju Paparan Radiasi, Multimeter X-Ray
Monitoring radiation exposure in each work area needs to be carried out to ensure that the radiation barriers, namely the walls in the radiation room, meet radiation safety requirements. This study aims to evaluate the value of the transmission factor for radiation protection in accordance with the provisions of NCRP 147, and to determine the results of measuring the rate of radiation exposure around the general radiography room at the Radiology Installation at RSUD dr. Zainoel Abidin Banda Aceh. The research method used was to calculate the thickness of the radiation barrier wall using the provisions of NCRP 147, then compare the results of the radiation barrier calculation with the thickness of the wall that has been built, then measure the rate of radiation exposure at the determined points. Calculation results of radiation barriers installed in the general radiography room at the Radiology Installation at RSUD dr. Zainoel Abidin Banda Aceh with 30 cm thick concrete and 2 mm Pb, for all walls meets the requirements and is safe for radiation workers and the general public. The results of calculating radiation exposure in the general radiography room at the Radiology Installation at RSUD dr. Zainoel Abidin Banda Aceh obtained a value that was far below the predetermined dose limit value so that the dose value was still in the safe category. Keywords: X-rays, Radiation Protection, Radiation Shielding, Radiation Exposure Rate, Multimeter X-Ray
SURVEI KESESUAIAN RUANG RADIOLOGI DI RUMAH SAKIT UMUM DAERAH LANGSA SEBAGAI UPAYA UNTUK PROTEKSI RADIASI (Putri Mawaddatul Muna, 2019)
PEMANTAUAN PROTEKSI RADIASI SINAR-X DI RUANG RADIOLOGI RUMAH SAKIT UMUM RUJUKAN PEMERINTAH ACEH UTARA (WAHYUDDIN, 2022)
PROTEKSI RADIASI RUANG RADIOLOGI DI UPTD RSUD DATU BERU KABUPATEN ACEH TENGAH (Rizke Dian Oktari, 2022)
ANALISIS PAPARAN RADIASI PESAWAT SINAR-X RONTGEN DI INSTALASI RADIOLOGI RUMAH SAKIT MEURAXA BANDA ACEH (Syahraicha, 2023)
UJI KESESUAIAN PESAWAT SINAR-X MEREK MINDRAY DI INSTALASI RUANG RADIOLOGI RUMAH SAKIT UMUM MEURAXA (Egidia Tesalonika, 2024)