ANALISIS DOSIS EFEKTIF PADA PEMERIKSAAN CT SCAN KEPALA DI RUMAH SAKIT TIPE A
Pengarang
Fadhillatul Husna - Personal Name;
Dosen Pembimbing
Rini Safitri - 197004251995122001 - Dosen Pembimbing I
Nomor Pokok Mahasiswa
1808102010019
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : Fakultas mipa., 2023
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Telah dilakukan analisis dosis efektif pada pemeriksaan CT Scan kepala di instalasi radiologi pada rumah sakit tipe A. Penelitian ini dilakukan dengan menggunakan data sekunder hasil pemeriksaan CT Scan kepala seperti nilai faktor eksposi, panjang scan, CTDIvol, dan DLP. Data dikelompokkan berdasarkan usia pasien yakni usia < 1 tahun, 1 – 4 tahun, 5 – 9 tahun, 10 - 17 tahun, dan > 18 tahun. Nilai dosis efektif yang dilakukan pada penelitian ini, yaitu nilai dosis efektif untuk bagian kepala yang didapatkan dari hasil perkalian nilai DLP dengan faktor bobot jaringan tubuh (k). Analisis dosis efektif dilakukan menggunakan nilai DRL standar peraturan Kepala BAPETEN dan ICRP 102. Dari hasil perhitungan diperoleh dosis efektif masing-masing kelompok yaitu pasien laki-laki < 1 tahun adalah 9,6 ± 2,4 mSv dan perempuan adalah 8 ± 3,3 mSv; usia 1 – 4 tahun, pasien laki-laki adalah 7,5 ± 2,1 mSv dan perempuan adalah 4,9 ± 0,7 mSv; usia 5 – 9 tahun, pasien laki-laki adalah 6,4 ± 3,4 mSv dan perempuan adalah 4,5 ± 1,6 mSv; usia 10 – 17 tahun, pasien laki-laki adalah 3,9 ± 0,8 mSv dan perempuan adalah 3,5 ± 0,5 mSv; usia > 18 tahun, pasien laki-laki adalah 3,9 ± 3,4 mSv dan perempuan adalah 2,4 ± 0,5 mSv. Nilai-nilai dosis yang diberikan masih sesuai dengan standar yang ditetapkan oleh BAPETEN dan ICRP, akan tetapi penyesuaian dosis DLP yang diberikan kepada pasien telah melebihi aturan yang telah ditentukan. Hal ini dikarenakan jumlah fase yang dilakukan pada setiap pemeriksaan berlebihan.
Kata Kunci : CT Scan, CTDIvol, DLP, Dosis Efektif, Kepala.
An effective dose analysis has been conducted on CT scan examination of the head at the radiology installation in a type A hospital. This study was conducted using secondary data of head CT scan examination results such as the value of exposure factor, scan length, CTDIvol, and DLP. The data were grouped based on the patient's age, namely age < 1 year, 1 - 4 years, 5 - 9 years, 10 - 17 years, and > 18 years. The effective dose value carried out in this study is the effective dose value for the head obtained from multiplying the DLP value by the body tissue weight factor (k). Effective dose analysis was carried out using the standard DRL value of the BAPETEN Head regulation and ICRP 102. From the calculation results, the effective dose of each group is obtained, namely male patients < 1 year is 9.6 ± 2.4 mSv and women are 8 ± 3.3 mSv; age 1 - 4 years, male patients are 7.5 ± 2.1 mSv and women are 4.9 ± 0.7 mSv; age 5 - 9 years, male patients were 6.4 ± 3.4 mSv and female were 4.5 ± 1.6 mSv; age 10 - 17 years, male patients were 3.9 ± 0.8 mSv and female were 3.5 ± 0.5 mSv; age > 18 years, male patients were 3.9 ± 3.4 mSv and female were 2.4 ± 0.5 mSv. The dose values given are still in accordance with the standards set by BAPETEN and ICRP, but the DLP dose adjustment given to patients has exceeded the predetermined rules. This is due to the of phases performed at each examination. Key Words : CT Scan, CTDIvol, DLP, Effective Dose, Head.
STUDI PENGARUH FAKTOR EKSPOSI, PANJANG SCAN, DAN PITCH TERHADAP DOSIS EFEKTIF PADA PEMERIKSAAN CT SCAN ABDOMEN KONTRAS (FAUZAN ABRARI, 2023)
PENGARUH SKOR CT SCAN HELSINKI DAN NILAI RASIO NEUTROFIL-LIMFOSIT TERHADAP LUARAN PASIEN CEDERA KEPALA (dr. Muhammad Ansari Adista, 2021)
PENGARUH DIAMETER EFEKTIF TERHADAP SIZE SPECIFIC DOSE ESTIMATE (SSDE) PADA PEMERIKSAAN CT-SCAN ABDOMEN DI RUMAH SAKIT UMUM DAERAH DR. ZAINOEL ABIDIN BANDA ACEH (Cintya Ikramina Tobing, 2024)
PENGUKURAN NILAI COMPUTER TOMOGRAPHY DOSE INDEX (CTDI) DAN DOSE LENGTH PRODUCT (DLP) PADA PEMERIKSAAN MULTISLICE COMPUTED TOMOGRAPHY (MSCT) DIAGNOSA KANKER PARU-PARU DI INSTALASI RADIOLOGI (Futri Nadia Vuspita, 2024)
ESTIMASI DOSIS RADIASI DAN FAKTOR RISIKO PADA PEMERIKSAAN CT SCAN THORAX DI RSUD DR. ZAINOEL ABIDIN DENGAN MEMBEDAKAN PENGGUNAAN MEDIA KONTRAS DANRNNON -KONTRAS (Faradilla Yusfa, 2023)