ANALISA SEBARAN DOSIS PAPARAN RADIASI PESAWAT SINAR-X SEBAGAI FUNGSI JARAK PADA LABORATORIUM KLINIK FAKULTAS KEDOKTERAN HEWAN UNIVERSITAS SYIAH KUALA | ELECTRONIC THESES AND DISSERTATION

Electronic Theses and Dissertation

Universitas Syiah Kuala

    SKRIPSI

ANALISA SEBARAN DOSIS PAPARAN RADIASI PESAWAT SINAR-X SEBAGAI FUNGSI JARAK PADA LABORATORIUM KLINIK FAKULTAS KEDOKTERAN HEWAN UNIVERSITAS SYIAH KUALA


Pengarang

Fitriani - Personal Name;

Dosen Pembimbing

Evi Yufita - 197509202000122001 - Dosen Pembimbing I
Rini Safitri - 197004251995122001 - Dosen Pembimbing II



Nomor Pokok Mahasiswa

1908102010061

Fakultas & Prodi

Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201

Subject
-
Kata Kunci
-
Penerbit

Banda Aceh : Fakultas MIPA Fisika., 2023

Bahasa

No Classification

-

Literature Searching Service

Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)

Analisa sebaran dosis paparan radiasi di ruang laboratorium klinik sangat penting untuk
mengetahui radiasi yang dihamburkan, sehingga dapat dipastikan jumlah radiasi yang
diterima oleh para pekerja radiasi. Tujuan penelitian ini adalah untuk mengetahui
pengaruh nilai tegangan tabung sinar-X terhadap besarnya dosis paparan radiasi di ruang
laboratorium klinik, mengetahui sebaran dosis paparan radiasi pada saat kegiatan
eksposi sebagai fungsi jarak dan mengetahui efektifitas perisai radiasi di laboratorium
klinik. Metode penelitian yang dilakukan adalah dengan melakukan pengukuran radiasi
primer langsung dibawah sumber radiasi sinar-X yaitu di titik pusat (0,0). Selanjutnya
pengukuran radiasi sekunder pada wilayah di sekitar sumber radiasi primer yaitu
pada jarak setiap 1 m, 1,5 m dan 2m dari sumber radiasi. Data efektifitas perisai radiasi
diukur pada bagian dalam dan luar perisai yaitu pintu, dinding dan kaca operator.
Berdasarkan hasil penelitian, bahwa semakin besar tegangan tabung sinar-X yang
diberikan maka semakin besar dosis radiasi yang dihasilkan. Pada tegangan46 kV,
arus 100 dan 200 mA menghasilkan dosis radiasinya sebesar 0,001334 – 0,003040 mGy.
Tegangan 55 kV, arus 100 dan 200 mA menghasilkan dosisradiasinya sebesar 0,002061-
0,005251 mGy dan tegangan 81 kV, arus 100 dan 200 mA menghasilkan dosis
radiasinya sebesar 0,006737- 0,009345 mGy. Berdasarkan pengukuran dosis sekunder
diperoleh bahwa sebaran dosis paparan radiasi pada saat eksposi sebagai fungsi jarak di
laboratorium klinik, pada pemberian tegangan tabung 46 kV, arus tabung 100 mA
dihasilkan nilai dosis radiasinya 0,000005 sd 0,001334 mGy dan untuk 200 mA
dihasilkan nilai dosis radiasinya 0,00001 sd 0,001693 mGy sehingga diperoleh area
aman yaitu lebih dari 1 meter dari sumber radiasi. Pada tegangan 55 kV, arus 100 mA
dihasilkan nilai dosis radiasinya 0,000005 sd 0,002618 mGy dan 200mA dihasilkan nilai
dosis radiasinya 0,000003 sd 0,004144 mGy sehingga diperoleh area aman pada jarak
1,5 meter dari sumber. Pada tegangan 81 kV, arus 100 mA dihasilkan nilai dosis
radiasinya 0,000009 sd 0,005673 mGy dan 200 mA, dihasilkan nilai dosis radiasinya
0,000014 sd 0,009737 mGy, sehingga area amannya berada pada jarak 2 meter dari
sumber radiasi. Efektifitas perisai radiasi di laboratorium klinik sangat baik (100 %),
pada pintu, dinding ruangan dan jendela kaca operator.

Analysis of the distribution of radiation exposure doses in clinical laboratory rooms is very important to determine the radiation scattered, so that it can be ascertained the amount of radiation received by radiation workers. The aim of this research is to determine the effect of the X-ray tube voltage value on the amount of radiation exposure dose in the clinical laboratory room, to determine the distribution of radiation exposure dose during exposure activities as a function of distance and to determine the effectiveness of radiation shielding in the clinical laboratory. The research method used is to measure primary radiation directly under the X-ray radiation source, namely at the center point (0.0). Furthermore, secondary radiation is radiated to the area around the primary radiation source, namely at a distance of 1 m, 1.5 m and 2 m from the radiation source. Radiation shield effectiveness data is measured on the inside and outside of the shield, namely doors, walls and operator glass. Based on research results, the greater the X-ray tube voltage applied, the greater the radiation dose produced. At a voltage of 46 kV, currents of 100 and 200 mA produce a radiation dose of 0.001334 – 0.003040 mGy. Voltage of 55 kV, current of 100 and 200 mA produces a radiation dose of 0.002061-0.005251 mGy and a voltage of 81 kV, current of 100 and 200 mA produces a radiation dose of 0.006737-0.009345 mGy. Based on secondary dose measurements, it was found that the distribution of radiation exposure dose during exposure as a function of distance in the clinical laboratory, when applying a tube voltage of 46 kV, a tube current of 100 mA produced a radiation dose value of 0.000005 to 0.001334 mGy and for 200 mA the resulting dose value The radiation is 0.00001 to 0.001693 mGy so that a safe area is obtained, namely more than 1 meter from the radiation source. At a voltage of 55 kV, a current of 100 mA produces a radiation dose value of 0.000005 to 0.002618 mGy and 200 mA produces a radiation dose value of 0.000003 to 0.004144 mGy so that a safe area is obtained at a distance of 1.5 meters from the source. At a voltage of 81 kV, a current of 100 mA produces a radiation dose value of 0.000009 to 0.005673 mGy and 200 mA produces a radiation dose value of 0.000014 to 0.009737 mGy, so that the safe area is at a distance of 2 meters from the radiation source. The effectiveness of radiation shielding in clinical laboratories is very good (100%), on doors, room walls and operator glass windows.