KOEFISIEN SELEKTIVITAS (KIJ) DAN PENGARUH TOTAL IONIC STRENGTH ADJUSTMENT BUFFER (TISAB) TERHADAP RESPON ESI AL3+
Pengarang
Firaihanil Jannah - Personal Name;
Dosen Pembimbing
Eka Safitri - 197001052000032001 - Dosen Pembimbing I
Nomor Pokok Mahasiswa
2008203010002
Fakultas & Prodi
Fakultas MIPA / Kimia (S2) / PDDIKTI : 47101
Subject
Kata Kunci
Penerbit
Banda Aceh : Fakultas MIPA Magister Kimia., 2023
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Pada penelitian ini telah berhasil dibuat elektroda selektif ion (ISE) untuk ion Al3+ menggunakan minyak jarak (Ricinus communis. L) berbasis membran poliuretan (PU) dengan bahan aktif 1,10-fenantrolin. Penelitian ini bertujuan untuk mempelajari pengaruh ion asing dan pengaruh Total Ionic Strength Adjustment Buffer (TISAB) terhadap respon ESI Al3+. Pengaruh ion asing diukur dengan menentukan koefisien selektivitas (Kij) menggunakan metode larutan terpisah. Larutan yang diukur yaitu ion utama Al3+ dan ion asing Na+, K+, Li+ , Pb2+, Ca2+, Co2+, Cu2+, Ni2+, Cr3+, dan Fe3+ dengan rentang konsentrasi 1x10-9 – 1x10-4 M. Hasil penelitian menunjukkan keberadaan ion asing tidak mempengaruhi kinerja ESI Al3+ dengan perolehan nilai Kij < 1. Urutan selektivitas ESI Al3+ yaitu Fe3+ > Cr3+ > Pb2+ > Cu2+ > Ni2+ > Co2+ > Ca2+ > Li+ > Na+ > K+. Pengaruh konsentrasi larutan TISAB diuji terhadap linieritas, rentang linier, LOD, dan LOQ. Hasil menunjukkan nilai sensitivitas yang tinggi pada konsentrasi TISAB 1x10-1, 1x10-2, 1x10-3, 1x10-4, dan 1x10-5 berturut-turut yaitu 18,50 ± 0,71, 18,52 ± 1,4, 18,18 ± 0,05, 18,89 ± 0,5 dan 19,15 ± 0,27. Respon ESI Al3+ yang diukur tanpa larutan TISAB mempunyai rentang linier yang lebih pendek yaitu 10-4 – 10-9 M, setelah dilakukan penambahan larutan TISAB rentang linier menjadi lebih lebar yaitu 10-4 – 10-10 M. Nilai LOD dan LOQ yang lebih rendah dengan penambahan TISAB, serta koefisien determinasi ≈ 1 yang berarti bahwa ESI Al3+ memiliki linieritas yang baik terhadap ion Al3+. Maka hasil dari penelitian ini diperoleh konsentrasi larutan TISAB terbaik adalah 1x10-5. Serta didapatkan nilai kij ESI Al3+ berbasis membran PU pada berbagai konsentrasi ion asing Na+, K+, Li+, Pb2+, Ca2+, Co2+, Cu2+, Ni2+, Cr3+, dan Fe3+.
In this research, an ion selective electrode (ISE) for Al3+ ions has been successfully developed using castor oil (Ricinus communis L) based on a polyurethane (PU) membrane with the active material 1,10-phenanthroline. The aim of this study was to investigate the influence of foreign ions and the Total Ionic Strength Adjustment Buffer (TISAB) on the response of the Al3+ ISE. The impact of foreign ions was evaluated by determining the selectivity coefficients (Kij) using the separate solution method. The measured solutions included the main ion Al3+ and the foreign ions Na+, K+, Li+ , Pb2+, Ca2+, Co2+, Cu2+, Ni2+, Cr3+, and Fe3+ over a concentration range of 1x10-9 – 1x10-4 M. The research results indicated that the presence of foreign ions did not affect the performance of the Al3+ ISE, as evidenced by Kij values < 1. The selectivity order of the Al3+ ISE was found to be Fe3+ > Cr3+ > Pb2+ > Cu2+ > Ni2+ > Co2+ > Ca2+ > Li+ > Na+ > K+. The influence of TISAB concentration on linearity, limit of detection (LOD), and limit of quantification (LOQ) were tested. The results showed high sensitivity at TISAB concentrations of 1x10-1, 1x10-2, 1x10-3, 1x10-4, and 1x10-5 were 18,50 ± 0,71, 18,52 ± 1,4, 18,18 ± 0,05, 18,89 ± 0,5, and 19,15 ± 0,27, respectively. The response of the Al3+ ISE measured without TISAB solution exhibited a shorter linear range of 10-4 – 10-9 M. After the addition of TISAB solution, the linear range extended to 10-4 – 10-10 M. Lower LOD and LOQ values were achieved with the addition of TISAB, and the coefficient of determination (≈ 1) indicated that the Al3+ ISE based on PU membrane exhibited good linearity towards Al3+ ions. Therefore, the optimal TISAB concentration obtained from this study was 1x10-5. Additionally, the Kij values of the PU membrane-based Al3+ ISE were determined at various concentrations of foreign ions including Na+, K+, Li+, Pb2+, Ca2+, Co2+, Cu2+, Ni2+, Cr3+, dan Fe3+.
PENENTUAN KOEFISIEN SELEKTIVITAS (KIJ) ELEKTRODA SELEKTIF ION (ESI) AL3+ BERBASIS MEMBRAN POLIURETAN TERMODIFIKASI GRAPHENE DAN 1,10-FENANTROLIN (REHAN AFIFAH HANJAYA, 2023)
PENGARUH KONSUMSI KEJU DAN SUSU TERHADAP KAPASITAS BUFFER SALIVA PADA MURID USIA 10-12 TAHUN DI SEKOLAH DASAR NEGERI 57 BANDA ACEH (HUSNUL KHATIMAH IMADUDDIN, 2017)
PENGARUH INFORMASI LABA TERHADAP KOEFISIEN RESPON LABA (STUDI EMPIRIS PADA PERUSAHAAN MANUFAKTUR YANG TERDAFTAR DI BEI PADA TAHUN 2011-2014) (NOVI LIYA SARI, 2016)
PENGARUH RISIKO SISTEMATIS TERHADAP PENGUNGKAPAN CORPORATE SOCIAL RESPONSIBILITY SERTA DAMPAKNYA TERHADAP PERTUMBUHAN LABA DAN KOEFISIEN RESPON LABA (STUD PADA PERUSAHAAN MANUFAKTUR YANG TERDAFTAR DI BURSA EFEK INDONESIA) (arief Andika, 2025)
ANALISIS PERBEDAAN KOEFISIEN RESPON LABA SEBELUM DAN SESUDAH PERUSAHAAN MEMBENTUK KOMITE AUDITRN(STUDI PADA PERUSAHAAN YANG TERDAFTAR DI BURSA EFEK INDONESIA) (Ramadhaniaty, 2024)