STUDI PENGARUH FAKTOR EKSPOSI, PANJANG SCAN, DAN PITCH TERHADAP DOSIS EFEKTIF PADA PEMERIKSAAN CT SCAN ABDOMEN KONTRAS
Pengarang
FAUZAN ABRARI - Personal Name;
Dosen Pembimbing
Elin Yusibani - 197903232006042002 - Dosen Pembimbing I
Zulfalina - 197507072002122001 - Dosen Pembimbing II
Nomor Pokok Mahasiswa
1808102010044
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : Fakultas MIPA Fisika., 2023
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Telah dilakukan penelitian di instalasi radiologi pada rumah sakit tipe A. Penelitian ini menggunakan data yang diperoleh berupa sampel dari hasil pemeriksaan CT Scan abdomen seperti nilai faktor eksposi, panjang scan, pitch, CTDIvol, dan DLP. Data akan di kelompokkan berdasarkan usia pasien yakni usia < 1 tahun, 1 – 4 tahun, 5 – 9 tahun, 10 - 17 tahun, dan ≥ 18 tahun. Nilai dosis efektif yang dilakukan pada penelitian ini, yaitu nilai dosis efektif untuk bagian tubuh abdomen yang didapatkan dari hasil perkalian nilai DLP dengan faktor bobot jaringan tubuh (k). Analisis dosis efektif dilakukan menggunakan nilai DRL standar peraturan Kepala BAPETEN dan ICRP 102. Dari hasil perhitungan diperoleh dosis efektif pada pasien CT Scan abdomen pada usia < 1 tahun adalah 8,5 ± 0,58 mSv; usia 1 – 4 tahun adalah 5,43 mSv; usia 5 – 9 tahun adalah 4,96 mSv; usia 10 – 17 tahun adalah 14,77 ± 0,95 mSv; usia ≥ 18 tahun adalah 20,57 ± 6,92 mSv. Kemudian pada pasien CT Scan abdomen + pelvis pada usia 5 – 9 tahun adalah 18,32 ± 3,5 mSv dan usia ≥ 18 tahun adalah 28,58 ± 6,58 mSv. Nilai-nilai dosis yang diberikan masih sesuai dengan standar yang ditetapkan oleh BAPETEN dan ICRP, akan tetapi penyesuaian dosis DLP yang diberikan kepada pasien telah melebihi aturan BAPETEN Nomor 1211 Tahun 2021 dan ICRP 102. Hal ini dikarenakan jumlah fase yang dilakukan pada setiap pemeriksaan berlebihan.
Kata Kunci: CT Scan, CTDIVol, DLP, Dosis Efektif, Abdomen, Kontras.
Research has been carried out in radiology installations at type A hospitals. The study was conducted by using secondary data from the CT Scan of the abdomen, namely the value of exposure factor, scan length, pitch, CTDIVol, and DLP. Data are grouped according to patients age namely age < 1 years, 1 – 4 years, 5 – 9 years, 10 – 17 years, ≥ 18 years. The effective dose value is obtained from the product of the DLP value with the conversion factor (k). Effective dose analysis was performed using standard DRL values of the Head of BAPETEN regulation and ICRP. From the calculation results, it is obtained that the effective dose in abdominal CT scan patients at the age of < 1 years were 8,5 ± 0,58 mSv; age 1- 4 years were 5,43 mSv; age 5 – 9 years were 4,96 mSv; age 10 – 17 years were 14,77 ± 0,95 mSv; age ≥ 18 years were 20,57 ± 6,92 mSv. then in abdominal + pelvic CT Scan patients at the age of 5 – 9 years were 18,32 ± 3,5 mSv, and age ≥ 18 years were 28,58 ± 6,58 mSv. The values of the doses given are still in accordance with the standards set by BAPETEN Number 1211 year 2021 and ICRP 102, but the adjustment of the DLP dose given to patients has exceeded the rules of BAPETEN and ICRP 102. This is because the number of phases carried out in each examination is excessive. Key Word: CT Scan, CTDIVol, DLP, Effective Dose, Abdominal, Kontras.
PENGUKURAN ENTRANCE SURFACE DOSE (ESD) TERHADAP RADIASI PESAWAT CT-SCAN PADA PEMERIKSAAN ABDOMEN DI RSUDZA BANDA ACEH (Nur Ulia Putri, 2022)
ESTIMASI DOSIS RADIASI DAN FAKTOR RISIKO PADA PEMERIKSAAN CT SCAN THORAX DI RSUD DR. ZAINOEL ABIDIN DENGAN MEMBEDAKAN PENGGUNAAN MEDIA KONTRAS DANRNNON -KONTRAS (Faradilla Yusfa, 2023)
ESTIMASI DOSIS RADIASI PADA PEMERIKSAAN CT- SCAN TORAKS BERDASARKAN FAKTOR EKSPOSI DAN KUALITAS CITRA MENGGUNAKAN PARAMETER PENILAIAN SIGNAL TO NOISE RATIO (SNR) (Ayunda Fajarni Rahmi, 2024)
PENGARUH DIAMETER EFEKTIF TERHADAP SIZE SPECIFIC DOSE ESTIMATE (SSDE) PADA PEMERIKSAAN CT-SCAN ABDOMEN DI RUMAH SAKIT UMUM DAERAH DR. ZAINOEL ABIDIN BANDA ACEH (Cintya Ikramina Tobing, 2024)
ANALISIS DOSIS EFEKTIF PADA PEMERIKSAAN CT SCAN KEPALA DI RUMAH SAKIT TIPE A (Fadhillatul Husna, 2023)