ANALISIS UJI KESESUAIAN GENERATOR DAN TABUNG PESAWAT SINAR-X DI RUANG RADIOLOGI RSUD ACEH BESAR
Pengarang
DITA SAFIRA - Personal Name;
Dosen Pembimbing
Rini Safitri - 197004251995122001 - Dosen Pembimbing I
Irhamni - 197101061997022001 - Dosen Pembimbing I
Nomor Pokok Mahasiswa
1808102010025
Fakultas & Prodi
Fakultas MIPA / Fisika (S1) / PDDIKTI : 45201
Subject
Kata Kunci
Penerbit
Banda Aceh : Fakultas mipa., 2023
Bahasa
No Classification
-
Literature Searching Service
Hard copy atau foto copy dari buku ini dapat diberikan dengan syarat ketentuan berlaku, jika berminat, silahkan hubungi via telegram (Chat Services LSS)
Rumah sakit Aceh Besar memiliki fasilitas ruang radiologi dengan mobile X-ray yang digunakan untuk keperluan diagnosa penyakit, dan alat tersebut telah berusia lebih dari sembilan tahun serta belum pernah dilakukannya uji kesesuaian. Uji kesesuaian adalah pengujian yang dilakukan untuk memastikan pesawat sinar-X dalam kondisi yang andal atau layak dan memenuhi peraturan perundang-undangan yang ditetapkan. Uji kesesuaian bertujuan untuk keamanan pasien, pekerja dan masyarakat dalam sistem pengoprasian pesawat sinar-X dengan baik dan benar. Uji yang dilakukan dalam penelitian ini adalah uji akurasi tegangan, uji linearitas dan uji kualitas berkas sinar-X (HVL). Alat yang digunakan adalah multimeter X-ray. Dalam pengukuran dibandingkan hasil yang terbaca pada alat dengan ketentuan yang telah ditetapkan BAPETEN no 9 tahun 2011 adalah uji akurasi tegangan ± ≤10%, uji linearitas ± ≤0.1 dan uji HVL ≥ 2.3 mmAl untuk 80kV. Hasil yang diperoleh pada uji akurasi tegangan yaitu nilai penyimpangan terbesar adalah 8.62% pada 46 kV, Uji linearitas yang diperoleh sebesar 0.014 dan pengukuran nilai kualitas sinar-X (HVL) adalah 2.6 mmAl pada 80 kV. Hasil yang dapat disimpulkan adalah pesawat sinar-X di ruang radiologi rumah sakit umum daerah Aceh Besar dalam kondisi yang andal sesuai Perka BAPETEN no 9 tahun 2011.
Kata kunci: Uji Kesesuaian, Sinar-X, Multimeter X-ray, mobile x-Ray
The Aceh Besar Hospital has a radiology room facility with a mobile X-ray that is used for disease diagnosis, but the device is more than nine years old and has never gotten a suitability test. It is a test conducted to ensure the X-ray machine is in a reliable or proper condition and meets the laws and regulations. The purpose of the suitability test is to ensure the safety of patients, workers, and the general public while the X-ray machine is in use. The tests that were carried out in this study were the accuracy of voltage, the linearity, and the X-ray beam quality (HVL). The study used an X-ray multimeter to compare the calculated value with the read value on the instrument. The BAPETEN Regulation No. 9 of 2011 states that the accuracy of voltage is ± ≤10%, the linearity is ± ≤0.1 and the HVL is ≥2.3 mmAl for 80kV. This research shows that the accuracy of voltage shows the largest deviation value of 8.62% at 46 kV, the linearity of 0.014, and the HVL of 2.6 mmAl at 80 kV. It is concluded that the X-ray machine in radiology at the Aceh Besar Hospital is in reliable condition and meets BAPETEN Regulation No. 9 of 2011. Keyword : Suitability test, X-Rays, X-ray multimeter, X-ray mobile
SURVEI KESESUAIAN RUANG RADIOLOGI DI RUMAH SAKIT UMUM DAERAH LANGSA SEBAGAI UPAYA UNTUK PROTEKSI RADIASI (Putri Mawaddatul Muna, 2019)
ANALISIS UJI KESESUAIAN GENERATOR DAN TABUNG PESAWAT SINAR-X DI RUANG RADIOLOGI RSUD ACEH BESAR (DITA SAFIRA, 2023)
PEMANTAUAN PROTEKSI RADIASI SINAR-X DI RUANG RADIOLOGI RUMAH SAKIT UMUM RUJUKAN PEMERINTAH ACEH UTARA (WAHYUDDIN, 2022)
QUALITY COTROL (KENDALI MUTU) PESAWAT SINAR-X MERK SIEMENS HELTINERS DI INSTALASI RADIOLOGI RUMAH SAKIT PENDIDIKAN UNIVERSITAS SYIAH KUALA (Mauliana, 2024)
PERHITUNGAN DAN ANALISIS DOSIS SERAP RADIASI FOTO THORAX PADA PASIEN ANAK (ULFA MAWARNI, 2022)